High-performance liquid chromatography is an analytical technique that separates components in a liquid sample. A pump moves a liquid mobile phase through a column packed with a solid stationary phase. Compounds interact differently with both phases and travel at different rates, leaving the column at distinct retention times. A detector records these arrivals as peaks on a chromatogram. The resulting pattern supports identification and quantification of substances in mixtures. Modern instruments use high pressure to force solvent through small particles, which improves speed and resolution compared with older low-pressure liquid chromatography methods.
Separation in HPLC depends on the chemistry of the stationary phase, the composition of the mobile phase, and the physical properties of the column. Reverse-phase separations use a nonpolar stationary phase and a polar mobile phase, and they are common for many organic compounds. Ion-exchange, size-exclusion, and normal-phase modes serve other classes of analytes. Gradient elution changes solvent strength over time, while isocratic elution holds it constant. Flow rate, temperature, particle size, and column length all influence peak shape and resolution. Detection may use ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry, depending on the analyte and the required sensitivity.
Routine HPLC testing compares a sample result with a calibration curve prepared from known reference standards. Peak area or peak height is plotted against concentration, and the curve is used to estimate unknown amounts. Retention time supports tentative identification when compared with a standard, though mass spectrometry or another confirmatory method may be needed for definitive identification. Pre-run checks verify repeatability, resolution, and peak symmetry before sample analysis. Limits of detection and quantification describe the smallest amounts that can be reliably observed or measured. Sample preparation, filtration, and degassing help prevent column damage and inconsistent results.
Routine quality control uses system suitability, blank injections, check standards, and control samples to detect drift or contamination. System suitability criteria may specify minimum resolution, maximum tailing factor, and a permitted range for repeated injections. Blank injections reveal carryover or solvent contamination, while check standards confirm calibration accuracy over a batch. Control samples with known analyte levels can show whether results remain within statistical limits. When a control result falls outside limits, the analyst investigates the cause and may invalidate affected results before repeating the batch.
Documentation and traceability are central to regulated HPLC testing. Records typically include instrument logs, column history, mobile-phase preparation, sample preparation, injection sequences, raw chromatograms, and audit trails. Electronic systems may require user access controls, time-stamped changes, and backup procedures. Training records show that analysts are qualified for assigned methods. Audits and inspections check whether written procedures match actual practice and whether deviations are documented. These controls support reproducibility and allow results to be reconstructed if questions arise later.
| Property | Value | Notes |
|---|---|---|
| Common abbreviation | HPLC | High-performance liquid chromatography |
| Separation basis | Differential partitioning | Between liquid mobile phase and solid stationary phase |
| Common mode | Reverse phase | Nonpolar column, polar mobile phase |
| Typical detector | UV-Vis absorbance | Widely used for compounds with chromophores |
| Typical column particle size | 2–5 µm | Smaller particles can improve resolution |
Method validation demonstrates that an HPLC procedure is suitable for its intended purpose. Common validation parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantification, and robustness. Accuracy reflects agreement with a reference value, while precision describes repeatability under defined conditions. Specificity shows whether the method can measure the analyte in the presence of impurities or matrix components. Validation documents are reviewed before a method is used for routine testing or regulatory submissions.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Typical checks include retention time, peak area precision, resolution between critical pairs, tailing factor, and theoretical plate count. Acceptance criteria are set in the method or pharmacopeial monograph. If a suitability check fails, the run may be rejected and the instrument or sample preparation may need investigation. This practice helps prevent release of data from a system that has drifted out of control.
Quality control samples are inserted at intervals to monitor accuracy and precision throughout a batch. Blank samples detect contamination, while spiked samples assess recovery from the sample matrix. Calibration standards establish the relationship between detector response and concentration, and control samples are prepared independently from them whenever possible. Laboratories also participate in proficiency testing and maintain audit trails, instrument logs, and reagent records. Ongoing review of control charts can reveal trends before they cause out-of-specification results.
Developing an HPLC method begins with defining the purpose, such as quantifying a main component, measuring impurities, or confirming identity. Analysts select separation mode, column, mobile phase, detection, and sample preparation based on analyte properties and matrix. Experiments vary solvent strength, pH, buffer type, and temperature to achieve resolution between critical peaks. The goal is a robust method that produces reliable results across instruments and operators. Method development often involves trial runs and statistical optimization.
Validation demonstrates that a method is suitable for its intended use. Typical performance characteristics include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantitation, and robustness. Regulators and standards organizations provide frameworks, but specific requirements depend on the application and jurisdiction. System suitability tests are run before sample analysis to confirm resolution, peak symmetry, retention time repeatability, and sensitivity. A validated method is not permanently fixed; changes may require partial or full revalidation.
Routine HPLC testing depends on controlled reagents, calibrated instruments, and documented procedures. Columns degrade over time, so retention times and peak shapes are monitored for drift. Mobile phases are filtered and degassed to prevent pump damage and detector noise. Reference standards must be traceable and stored under suitable conditions. Data handling systems record injections, calculations, and audit trails. Quality control samples interspersed with unknowns help detect errors during a run.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Common checks include retention time, peak area, resolution between critical pairs, tailing factor, and theoretical plate count. Results are compared with predefined limits, and a failed check requires investigation before sample results are reported. Quality control samples at low, middle, and high concentrations are injected at intervals to monitor accuracy and precision. Blank injections detect carryover and contamination, while control charts track performance over time.
Data handling and documentation are central to HPLC quality control. Electronic systems should have audit trails that record changes to methods, sequences, and results. Integration parameters, such as peak baseline and threshold, can affect reported areas and must be defined in advance. Out-of-specification results trigger a structured investigation that may include reanalysis, instrument checks, and review of sample preparation. Regulatory inspections often examine raw data, audit trails, and training records to verify that reported results are traceable and reliable.
Mexico (south-eastern): Among the ancient Maya, Nicotiana was considered a sacred plant, closely associated with deities of earth and sky, and used for both visionary and therapeutic ends. The contemporary Tzeltal and Tzotzil Maya of Highland Chiapas (Mexico) are bearers of this ethnobotanical inheritance, preserving a rich and varied tradition of Nicotiana use and folklore. The entire tobacco plant is viewed as a primordial medicine and a powerful botanical "helper" or "protector". Depending on the condition to be treated, whole Nicotiana leaves are used alone or in combination with other herbs in the preparation of various medicinal plasters and teas. In its most common form, fresh or green leaves are ground with slaked lime to produce an intoxicating oral snuff that serves as both a protective and therapeutic agent. United States: Extract of N. tabacum is taken orally to treat tiredness, ward off diseases, and quiet fear. Tanzania: Leaves of Nicotiana tabacum are placed in the vagina to stimulate labor. Zimbabwe: Leaves or root of the plant are infused and taken by mouth for asthma and other respiratory problems. Leaves and roots are also rubbed against warts and wounds as ointment.
== Medical uses == Penbutolol is used to treat mild to moderate high blood pressure. Like other beta blockers it is not a first line treatment for this indication. It should not be used or only used with caution in people with heart failure and people with asthma. It may mask signs of low blood sugar in people with diabetes and it may mask signs of hyperthyroidism. Animal studies showed some signs of potential trouble for women who are pregnant, and it has not been tested in women who are pregnant. It is not known if penbutolol is secreted in breast milk.
They show that levels of these so-called "forever chemicals" in rainwater are ubiquitously, and often greatly, above guideline safe levels worldwide. There are moves to restrict and replace their use.On 18 August, a simple method of breaking down these chemicals once they have been pulled out of contaminated water or soil is described. A preprint describes the donations-funded plans of The Galileo Project in detail: a systematic scientific research program searching for (signs of) extraterrestrial technological civilizations (ETCs) on and near Earth. 3 August – Scientists report an organ perfusion system that can restore, i.e. on the cellular level, multiple vital (pig) organs one hour after death (during which the body had warm ischaemia), after reporting a similar method/system for reviving (pig) brains hours after death in 2019. This could be used to preserve donor organs or for revival in medical emergencies. 4 August – Lab-made cartilage gel based on a synthetic hydrogel composite is found to have greater strength and wear resistance than natural cartilage, which could enable the durable resurfacing of damaged articulating joints. 8 August Researchers provide a dataset of standardized calculated detailed environmental impacts of >57,000 circulating food products, potentially e.g. informing consumers or policy. A study quantifies the large extent of climate change impacts on infectious diseases.
The APPI interface for LC–MS was developed simultaneously by Bruins and Syage in 2000. APPI is another LC–MS ion source/ interface for the analysis of neutral compounds that cannot be ionized using ESI. This interface is similar to the APCI ion source, but instead of a corona discharge, the ionization occurs by using photons coming from a discharge lamp. In the direct-APPI mode, singly charged analyte molecular ions are formed by absorption of a photon and ejection of an electron. In the dopant-APPI mode, an easily ionizable compound (dopant) is added to the mobile phase or the nebulizing gas to promote a reaction of charge-exchange between the dopant molecular ion and the analyte. The ionized sample is later transferred to the mass analyzer at high vacuum as it passes through small orifice skimmers.
Sources: en.wikipedia.org
Comparatively simple devices are often used to apply counterions of increasing gradient to a chromatography column. Counterions such as copper (II) are chosen most often for effectively separating peptides and amino acids through complex formation. A simple device can be used to create a salt gradient. Elution buffer is consistently being drawn from the chamber into the mixing chamber, thereby altering its buffer concentration. Generally, the buffer placed into the chamber is usually of high initial concentration, whereas the buffer placed into the stirred chamber is usually of low concentration. As the high concentration buffer from the left chamber is mixed and drawn into the column, the buffer concentration of the stirred column gradually increase. Altering the shapes of the stirred chamber, as well as of the limit buffer, allows for the production of concave, linear, or convex gradients of counterion. A multitude of different mediums are used for the stationary phase. Among the most common immobilized charged groups used are trimethylaminoethyl (TAM), triethylaminoethyl (TEAE), diethyl-2-hydroxypropylaminoethyl (QAE), aminoethyl (AE), diethylaminoethyl (DEAE), sulpho (S), sulphomethyl (SM), sulphopropyl (SP), carboxy (C), and carboxymethyl (CM). Successful packing of the column is an important aspect of ion chromatography. Stability and efficiency of a final column depends on packing methods, solvent used, and factors that affect mechanical properties of the column.
== Identification of the mould == After structural comparison with different species of Penicillium, Fleming initially believed that his specimen was Penicillium chrysogenum, a species described by an American microbiologist Charles Thom in 1910. He was fortunate that Charles John Patrick La Touche, an Irish botanist, had just recently joined St Mary's as a mycologist to investigate fungi as the cause of asthma. La Touche identified the specimen as Penicillium rubrum, the identification used by Fleming in his publication. In 1931, Thom re-examined different Penicillium including that of Fleming's specimen. He came to a confusing conclusion, stating, "Ad. 35 [Fleming's specimen] is P. notatum WESTLING. This is a member of the P. chrysogenum series with smaller conidia than P. chrysogenum itself." From then on, Fleming's mould was synonymously referred to as P. notatum and P. chrysogenum. But Thom adopted and popularised the use of P. chrysogenum. In addition to P. notatum, newly discovered species such as P. meleagrinum and P. cyaneofulvum were recognised as members of P. chrysogenum in 1977. To resolve the confusion, the Seventeenth International Botanical Congress held in Vienna, Austria, in 2005 formally adopted the name P. chrysogenum as the conserved name (nomen conservandum). Whole genome sequence and phylogenetic analysis in 2011 revealed that Fleming's mould belongs to P. rubens, a species described by Belgian microbiologist Philibert Biourge in 1923.
== Quarks and the mass of a proton == In quantum chromodynamics, the modern theory of the nuclear force, most of the mass of protons and neutrons is explained by special relativity. The mass of a proton is about 80–100 times greater than the sum of the rest masses of its three valence quarks, while the gluons have zero rest mass. The extra energy of the quarks and gluons in a proton, as compared to the rest energy of the quarks alone in the QCD vacuum, accounts for almost 99% of the proton's mass. The rest mass of a proton is, thus, the invariant mass of the system of moving quarks and gluons that make up the particle, and, in such systems, even the energy of massless particles confined to a system is still measured as part of the rest mass of the system. Two terms are used in referring to the mass of the quarks that make up protons: current quark mass refers to the mass of a quark by itself, while constituent quark mass refers to the current quark mass plus the mass of the gluon particle field surrounding the quark. These masses typically have very different values. The kinetic energy of the quarks that is a consequence of confinement is a contribution (see Mass in special relativity). Using lattice QCD calculations, the contributions to the mass of the proton are the quark condensate (~9%, comprising the up and down quarks and a sea of virtual strange quarks), the quark kinetic energy (~32%), the gluon kinetic energy (~37%), and the anomalous gluonic contribution (~23%, comprising contributions from condensates of all quark flavors).
Sources: en.wikipedia.org
=== Prenatal diagnostic testing === The preferred prenatal diagnostic method is molecular analysis which requires a mutational analysis to measure BCKAD (branched-chain alpha-keto acid dehydrogenase) enzyme activity in chorion villus cells or amniocytes. Another measurement method is BCAA (branched-chain amino acid) concentration in amniotic fluids.
Galen believed that the arterial blood was created by venous blood passing from the left ventricle to the right by passing through 'pores' in the interventricular septum, air passed from the lungs via the pulmonary artery to the left side of the heart. As the arterial blood was created 'sooty' vapors were created and passed to the lungs also via the pulmonary artery to be exhaled. In addition, Ibn al-Nafis had an insight into what would become a larger theory of the capillary circulation. He stated that "there must be small communications or pores (manafidh in Arabic) between the pulmonary artery and vein," a prediction that preceded the discovery of the capillary system by more than 400 years. Ibn al-Nafis' theory, however, was confined to blood transit in the lungs and did not extend to the entire body. Finally, William Harvey, a pupil of Hieronymus Fabricius (who had earlier described the valves of the veins without recognizing their function), performed a sequence of experiments, and published Exercitatio Anatomica de Motu Cordis et Sanguinis in Animalibus in 1628, which "demonstrated that there had to be a direct connection between the venous and arterial systems throughout the body, and not just the lungs. Most importantly, he argued that the beat of the heart produced a continuous circulation of blood through minute connections at the extremities of the body.
Nelumbo nucifera is a lotus species with historical, cultural and spiritual significance. It is a sacred flower in both Hinduism and Buddhism, representing the path to spiritual awakening and enlightenment. Lotus is closely associated with Goddess Lakshmi and her consort Vishnu. In Hindu iconography, Lakshmi is shown either in seated posture or as standing on top of a lotus flower and holding lotuses in two of her four hands. Vishnu is always portrayed with a lotus in one of his hands. Lotus or Padma exemplifies the path of one who leads a dharmic life. One can stay in the darkness, trapped in the comfort of the decaying leaves and murky water or struggle to lead a dharmic life upholding righteousness, following the light and eventually liberate oneself of his/her past thereby leading a prosperous peaceful happy life. Lotus also signifies moksha. The roots of lotus plant intertwined in mud representing the cycle of life and death and the lotus flower symbolizing the serene blissful state of moksha. Lotus is also used to depict the awakening of Kundalini energy in the tantric system of Hinduism. In Hindu literature, lotus petals are used as a simile to praise the eyes of gods and goddesses. For instance, the word Kamalanayana (lotus-eyed) refers to Vishnu and it extols his eyes and praise him for being the one who holds the gaze/adoration of Kamala (Lakshmi).
Sources: en.wikipedia.org
HPLC testing measures the presence and amount of one or more compounds in a liquid sample. It separates mixture components and records detector responses as peaks, which are compared with reference standards. Results are usually reported as concentrations or relative percentages.
Retention time is the interval between sample injection and the detector response for a given compound. It depends on the compound's interactions with the stationary and mobile phases under set conditions. Matching a retention time to a standard supports tentative identification but is not always unique.
HPLC alone can separate unknown compounds and provide retention times, but it often cannot identify them with certainty. Coupling HPLC to mass spectrometry gives mass information that improves identification. Confirmation usually requires comparison with reference standards or complementary techniques.
Validation establishes suitability for a new method, while verification confirms that a method works in a specific laboratory. Verification is often used when a validated method is adopted with existing equipment and staff. Both rely on documented acceptance criteria.